09-24-2021-1005 - Absorption spectroscopy

Absorption spectroscopy refers to spectroscopic techniques that measure the absorption of radiation, as a function of frequency or wavelength, due to its interaction with a sample. The sample absorbs energy, i.e., photons, from the radiating field. The intensity of the absorption varies as a function of frequency, and this variation is the absorption spectrum. Absorption spectroscopy is performed across the electromagnetic spectrum.

Absorption spectroscopy is employed as an analytical chemistry tool to determine the presence of a particular substance in a sample and, in many cases, to quantify the amount of the substance present. Infrared and ultraviolet–visible spectroscopy are particularly common in analytical applications. Absorption spectroscopy is also employed in studies of molecular and atomic physics, astronomical spectroscopy and remote sensing.

There is a wide range of experimental approaches for measuring absorption spectra. The most common arrangement is to direct a generated beam of radiation at a sample and detect the intensity of the radiation that passes through it. The transmitted energy can be used to calculate the absorption. The source, sample arrangement and detection technique vary significantly depending on the frequency range and the purpose of the experiment.

Following are the major types of absorption spectroscopy:[1]

Sr. NoElectromagnetic RadiationSpectroscopic type
1X-rayX-ray absorption spectroscopy
2Ultraviolet–visibleUV–vis absorption spectroscopy
3InfraredIR absorption spectroscopy
4MicrowaveMicrowave absorption spectroscopy
5Radio waveElectron spin resonance spectroscopy

Nuclear magnetic resonance spectroscopy


Absorption spectrum[edit]

Solar spectrum with Fraunhofer lines as it appears visually

A material's absorption spectrum is the fraction of incident radiation absorbed by the material over a range of frequencies of Electromagnetic Radiation. The absorption spectrum is primarily determined[2][3][4] by the atomic and molecular composition of the material. Radiation is more likely to be absorbed at frequencies that match the energy difference between two quantum mechanical states of the molecules. The absorption that occurs due to a transition between two states is referred to as an absorption line and a spectrum is typically composed of many lines.

The frequencies where absorption lines occur, as well as their relative intensities, primarily depend on the electronic and molecular structure of the sample. The frequencies will also depend on the interactions between molecules in the sample, the crystal structure in solids, and on several environmental factors (e.g., temperature, pressure, electromagnetic field). The lines will also have a width and shape that are primarily determined by the spectral density or the density of states of the system.

Theory[edit]

Absorption lines are typically classified by the nature of the quantum mechanical change induced in the molecule or atom. Rotational lines, for instance, occur when the rotational state of a molecule is changed. Rotational lines are typically found in the microwave spectral region. Vibrational lines correspond to changes in the vibrational state of the molecule and are typically found in the infrared region. Electronic lines correspond to a change in the electronic state of an atom or molecule and are typically found in the visible and ultraviolet region. X-ray absorptions are associated with the excitation of inner shellelectrons in atoms. These changes can also be combined (e.g. rotation-vibration transitions), leading to new absorption lines at the combined energy of the two changes.

The energy associated with the quantum mechanical change primarily determines the frequency of the absorption line but the frequency can be shifted by several types of interactions. Electric and magnetic fields can cause a shift. Interactions with neighboring molecules can cause shifts. For instance, absorption lines of the gas phase molecule can shift significantly when that molecule is in a liquid or solid phase and interacting more strongly with neighboring molecules.

The width and shape of absorption lines are determined by the instrument used for the observation, the material absorbing the radiation and the physical environment of that material. It is common for lines to have the shape of a Gaussian or Lorentzian distribution. It is also common for a line to be described solely by its intensity and width instead of the entire shape being characterized.

The integrated intensity—obtained by integrating the area under the absorption line—is proportional to the amount of the absorbing substance present. The intensity is also related to the temperature of the substance and the quantum mechanical interaction between the radiation and the absorber. This interaction is quantified by the transition moment and depends on the particular lower state the transition starts from, and the upper state it is connected to.

The width of absorption lines may be determined by the spectrometer used to record it. A spectrometer has an inherent limit on how narrow a line it can resolve and so the observed width may be at this limit. If the width is larger than the resolution limit, then it is primarily determined by the environment of the absorber. A liquid or solid absorber, in which neighboring molecules strongly interact with one another, tends to have broader absorption lines than a gas. Increasing the temperature or pressure of the absorbing material will also tend to increase the line width. It is also common for several neighboring transitions to be close enough to one another that their lines overlap and the resulting overall line is therefore broader yet.

Relation to transmission spectrum[edit]

Absorption and transmission spectra represent equivalent information and one can be calculated from the other through a mathematical transformation. A transmission spectrum will have its maximum intensities at wavelengths where the absorption is weakest because more light is transmitted through the sample. An absorption spectrum will have its maximum intensities at wavelengths where the absorption is strongest.

Relation to emission spectrum[edit]

The emission spectrum of iron

Emission is a process by which a substance releases energy in the form of electromagnetic radiation. Emission can occur at any frequency at which absorption can occur, and this allows the absorption lines to be determined from an emission spectrum. The emission spectrum will typically have a quite different intensity pattern from the absorption spectrum, though, so the two are not equivalent. The absorption spectrum can be calculated from the emission spectrum using Einstein coefficients.

Relation to scattering and reflection spectra[edit]

The scattering and reflection spectra of a material are influenced by both its index of refraction and its absorption spectrum. In an optical context, the absorption spectrum is typically quantified by the extinction coefficient, and the extinction and index coefficients are quantitatively related through the Kramers-Kronig relation. Therefore, the absorption spectrum can be derived from a scattering or reflection spectrum. This typically requires simplifying assumptions or models, and so the derived absorption spectrum is an approximation.

Applications[edit]

The infrared absorption spectrum of NASA laboratory sulfur dioxide ice is compared with the infrared absorption spectra of ices on Jupiter's moon, Iocredit NASA, Bernard Schmitt, and UKIRT.

Absorption spectroscopy is useful in chemical analysis[5] because of its specificity and its quantitative nature. The specificity of absorption spectra allows compounds to be distinguished from one another in a mixture, making absorption spectroscopy useful in wide variety of applications. For instance, Infrared gas analyzers can be used to identify the presence of pollutants in the air, distinguishing the pollutant from nitrogen, oxygen, water and other expected constituents.[6]

The specificity also allows unknown samples to be identified by comparing a measured spectrum with a library of reference spectra. In many cases, it is possible to determine qualitative information about a sample even if it is not in a library. Infrared spectra, for instance, have characteristics absorption bands that indicate if carbon-hydrogen or carbon-oxygen bonds are present.

An absorption spectrum can be quantitatively related to the amount of material present using the Beer–Lambert law. Determining the absolute concentration of a compound requires knowledge of the compound's absorption coefficient. The absorption coefficient for some compounds is available from reference sources, and it can also be determined by measuring the spectrum of a calibration standard with a known concentration of the target.

Remote sensing[edit]

One of the unique advantages of spectroscopy as an analytical technique is that measurements can be made without bringing the instrument and sample into contact. Radiation that travels between a sample and an instrument will contain the spectral information, so the measurement can be made remotely. Remote spectral sensing is valuable in many situations. For example, measurements can be made in toxic or hazardous environments without placing an operator or instrument at risk. Also, sample material does not have to be brought into contact with the instrument—preventing possible cross contamination.

Remote spectral measurements present several challenges compared to laboratory measurements. The space in between the sample of interest and the instrument may also have spectral absorptions. These absorptions can mask or confound the absorption spectrum of the sample. These background interferences may also vary over time. The source of radiation in remote measurements is often an environmental source, such as sunlight or the thermal radiation from a warm object, and this makes it necessary to distinguish spectral absorption from changes in the source spectrum.

To simplify these challenges, Differential optical absorption spectroscopy has gained some popularity, as it focusses on differential absorption features and omits broad-band absorption such as aerosol extinction and extinction due to rayleigh scattering. This method is applied to ground-based, air-borne and satellite based measurements. Some ground-based methods provide the possibility to retrieve tropospheric and stratospheric trace gas profiles.

Astronomy[edit]

Absorption spectrum observed by the Hubble Space Telescope

Astronomical spectroscopy is a particularly significant type of remote spectral sensing. In this case, the objects and samples of interest are so distant from earth that electromagnetic radiation is the only means available to measure them. Astronomical spectra contain both absorption and emission spectral information. Absorption spectroscopy has been particularly important for understanding interstellar clouds and determining that some of them contain molecules. Absorption spectroscopy is also employed in the study of extrasolar planets. Detection of extrasolar planets by the transit method also measures their absorption spectrum and allows for the determination of the planet's atmospheric composition,[7] temperature, pressure, and scale height, and hence allows also for the determination of the planet's mass.[8]

Atomic and molecular physics[edit]

Theoretical models, principally quantum mechanical models, allow for the absorption spectra of atoms and molecules to be related to other physical properties such as electronic structure, atomic or molecular mass, and molecular geometry. Therefore, measurements of the absorption spectrum are used to determine these other properties. Microwave spectroscopy, for example, allows for the determination of bond lengths and angles with high precision.

In addition, spectral measurements can be used to determine the accuracy of theoretical predictions. For example, the Lamb shift measured in the hydrogenatomic absorption spectrum was not expected to exist at the time it was measured. Its discovery spurred and guided the development of quantum electrodynamics, and measurements of the Lamb shift are now used to determine the fine-structure constant.

Experimental methods[edit]

Basic approach[edit]

The most straightforward approach to absorption spectroscopy is to generate radiation with a source, measure a reference spectrum of that radiation with a detector and then re-measure the sample spectrum after placing the material of interest in between the source and detector. The two measured spectra can then be combined to determine the material's absorption spectrum. The sample spectrum alone is not sufficient to determine the absorption spectrum because it will be affected by the experimental conditions—the spectrum of the source, the absorption spectra of other materials in between the source and detector and the wavelength dependent characteristics of the detector. The reference spectrum will be affected in the same way, though, by these experimental conditions and therefore the combination yields the absorption spectrum of the material alone.

A wide variety of radiation sources are employed in order to cover the electromagnetic spectrum. For spectroscopy, it is generally desirable for a source to cover a broad swath of wavelengths in order to measure a broad region of the absorption spectrum. Some sources inherently emit a broad spectrum. Examples of these include globars or other black body sources in the infrared, mercury lamps in the visible and ultraviolet and x-ray tubes. One recently developed, novel source of broad spectrum radiation is synchrotron radiation which covers all of these spectral regions. Other radiation sources generate a narrow spectrum but the emission wavelength can be tuned to cover a spectral range. Examples of these include klystrons in the microwave region and lasers across the infrared, visible and ultraviolet region (though not all lasers have tunable wavelengths).

The detector employed to measure the radiation power will also depend on the wavelength range of interest. Most detectors are sensitive to a fairly broad spectral range and the sensor selected will often depend more on the sensitivity and noise requirements of a given measurement. Examples of detectors common in spectroscopy include heterodyne receivers in the microwave, bolometers in the millimeter-wave and infrared, mercury cadmium telluride and other cooled semiconductor detectors in the infrared, and photodiodes and photomultiplier tubes in the visible and ultraviolet.

If both the source and the detector cover a broad spectral region, then it is also necessary to introduce a means of resolving the wavelength of the radiation in order to determine the spectrum. Often a spectrograph is used to spatially separate the wavelengths of radiation so that the power at each wavelength can be measured independently. It is also common to employ interferometry to determine the spectrum—Fourier transform infrared spectroscopy is a widely used implementation of this technique.

Two other issues that must be considered in setting up an absorption spectroscopy experiment include the optics used to direct the radiation and the means of holding or containing the sample material (called a cuvette or cell). For most UV, visible, and NIR measurements the use of precision quartz cuvettes are necessary. In both cases, it is important to select materials that have relatively little absorption of their own in the wavelength range of interest. The absorption of other materials could interfere with or mask the absorption from the sample. For instance, in several wavelength ranges it is necessary to measure the sample under vacuum or in a rare gas environment because gases in the atmosphere have interfering absorption features.

Specific approaches[edit]

See also[edit]


https://en.wikipedia.org/wiki/Absorption_spectroscopy

 


09-24-2021-1003 - Linear elasticity and Linear Map

Ibiza Light 30cm Mirrorball 30mm Mirror Ball Revolving Glitter Ball Light  Effect 80's - SKU8852 | Simply Sound & Lighting

 Linear elasticity is a mathematical model of how solid objects deform and become internally stressed due to prescribed loading conditions. It is a simplification of the more general nonlinear theory of elasticity and a branch of continuum mechanics.

The fundamental "linearizing" assumptions of linear elasticity are: infinitesimal strains or "small" deformations(or strains) and linear relationships between the components of stress and strain. In addition linear elasticity is valid only for stress states that do not produce yielding.

These assumptions are reasonable for many engineering materials and engineering design scenarios. Linear elasticity is therefore used extensively in structural analysis and engineering design, often with the aid of finite element analysis.

https://en.wikipedia.org/wiki/Linear_elasticity

https://en.wikipedia.org/wiki/Linear_map



https://en.wikipedia.org/wiki/Two-photon_absorption

https://en.wikipedia.org/wiki/Nonlinear_optics

09-24-2021-1002 - Apodization

 Apodization is an optical filtering technique. Its literal translation from Greek is "removing the foot". It is the technical term for changing the shape of a mathematical function, an electrical signal, an optical transmission or a mechanical structure. In optics, it is primarily used to remove Airy disks caused by diffraction around an intensity peak, improving the focus.

Airy disk

Apodization in electronics[edit]

Apodization in signal processing[edit]

The term apodization is used frequently in publications on Fourier-transform infrared (FTIR) signal processing. An example of apodization is the use of the Hann window in the fast Fourier transform analyzer to smooth the discontinuities at the beginning and end of the sampled time record.

Apodization in digital audio[edit]

An apodizing filter can be used in digital audio processing instead of the more common brickwall filters, in order to avoid the pre-ringing that the latter introduces.

Apodization in mass spectrometry[edit]

During oscillation within an Orbitrap, ion transient signal may not be stable until the ions settle into their oscillations. Toward the end, subtle ion collisions have added up to cause noticeable dephasing. This presents a problem for the Fourier transformation, as it averages the oscillatory signal across the length of the time-domain measurement. Software allows “apodization”, the removal of the front and back section of the transient signal from consideration in the FT calculation. Thus, apodization improves the resolution of the resulting mass spectrum. Another way to improve the quality of the transient is to wait to collect data until ions have settled into stable oscillatory motion within the trap.[1]

Apodization in optics[edit]

In optical design jargon, an apodization function is used to purposely change the input intensity profile of an optical system, and may be a complicated function to tailor the system to certain properties. Usually it refers to a non-uniform illumination or transmission profile that approaches zero at the edges.

Apodization in imaging[edit]

Since side lobes of the Airy disk are responsible for degrading the image, techniques for suppressing them are utilized. In case the imaging beam has Gaussian distribution, when the truncation ratio (the ratio of the diameter of the Gaussian beam to the diameter of the truncating aperture) is set to 1, the side-lobes become negligible and the beam profile becomes purely Gaussian. [2][page needed]

Apodization in photography[edit]

Most camera lenses contain diaphragms which decrease the amount of light coming into the camera. These are not strictly an example of apodization, since the diaphragm does not produce a smooth transition to zero intensity, nor does it provide shaping of the intensity profile (beyond the obvious all-or-nothing, "top hat" transmission of its aperture). 

Some lenses use other methods to reduce the amount of light let in. For example, the Minolta/Sony STF 135mm f/2.8 T4.5 lens however, has a special design introduced in 1999, which accomplishes this by utilizing a concave neutral-gray tinted lens element as an apodization filter, thereby producing a pleasant bokeh. The same optical effect can be achieved combining depth-of-field bracketing with multi exposure, as implemented in the Minolta Maxxum 7's STF function. In 2014, Fujifilm announced a lens utilizing a similar apodization filter in the Fujinon XF 56mm F1.2 R APD lens.[3] In 2017, Sonyintroduced the E-mount full-frame lens Sony FE 100mm F2.8 STF GM OSS (SEL-100F28GM) based on the same optical Smooth Trans Focus principle.[4]

Simulation of a Gaussian laser beam input profile is also an example of apodization.[citation needed]

Photon sieves provide a relatively easy way to achieve tailored optical apodization.[5]

Apodization in astronomy[edit]

Apodization is used in telescope optics in order to improve the dynamic range of the image. For example, stars with low intensity in the close vicinity of very bright stars can be made visible using this technique, and even images of planets can be obtained when otherwise obscured by the bright atmosphere of the star they orbit.[6][7][8] Generally, apodization reduces the resolution of an optical image; however, because it reduces diffraction edge effects, it can actually enhance certain small details. In fact the notion of resolution, as it is commonly defined with the Rayleigh criterion, is in this case partially irrelevant. One has to understand that the image formed in the focal plane of a lens (or a mirror) is modelled through the Fresnel diffraction formalism. The classical diffraction pattern, the Airy disk, is connected to a circular pupil, without any obstruction and with a uniform transmission. Any change in the shape of the pupil (for example a square instead of a circle), or in its transmission, results in an alteration in the associated diffraction pattern.

See also[edit]

References[edit]

  1. ^ Savaryn, John P.; Toby, Timothy K.; Kelleher, Neil L. (September 2016). "A researcher's guide to mass spectrometry-based proteomics". Proteomics. 16 (18): 2435–2443. doi:10.1002/pmic.201600113. PMC 5198776. PMID 27553853.
  2. ^ Handbook of optical and laser scanning. Marshall, Gerald F., Stutz, Glenn E. (2nd ed.). Boca Raton, Florida: CRC Press. 2012. ISBN 9781439808795. OCLC 756724023.
  3. ^ ""Bokeh-Gigant": Fujinon XF 1,2/56 mm R APD (aktualisiert)". 2001-11-30.
  4. ^ "Neu von Sony: E-Mount-Objektive 100 mm F2.8 STF GM, FE 85 mm F1.8; Blitz HVL-F45RM". Photoscala (in German). 2017-02-07. Archived from the original on 2017-02-11. Retrieved 2017-02-10.
  5. ^ Hewett, Jacqueline (2007-06-01). "Photon sieves benefit space telescopes". Optics.org. Retrieved 2007-06-05.
  6. ^ E. Hecht (1987). Optics (2nd ed.). Addison Wesley. ISBN 978-0-201-11609-0. Section 11.3.3.
  7. ^ FIRST RESULTS FROM VERY LARGE TELESCOPE NACO APODIZING PHASE PLATE: 4 μm IMAGES OF THE EXOPLANET β PICTORIS b* The Astrophysical Journal (Letter)
  8. ^ Planet hunters no longer blinded by the light. spacefellowship.com Note: this article includes several images of such a phase plate

https://en.wikipedia.org/wiki/Apodization